DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 124 Chapter 6 - Table
6.2.5.4 Remove the Pedal Assembly
1.) Refer to Figure 6-15. Use a 6 mm Allen wrench to remove the two socket-head cap screws
that fasten the pedal to the side of the table base.
Figure 6-15: Remove Pedal from Table Rail
2.) Remove the front left table side cover from the scissor section of the table.
a.) Use an 1/8 inch Allen wrench to remove the two screws and ground wire that fasten the
front left table side cover into place. (Only need to remove one cover.)
b.) If necessary, move the table to access the front switch rail screws.
6.2.5.5 Remove the PET and CT Position Switches and Switch Rail
Locate the CT and PET position limit switches (mounted to the switch rail and attached to the left
side of the secondary base when facing the gantry). Move the table, as needed, to access the
Position switch hardware and components. The cables may be cut and the switches and switch rail
removed as a single assembly.
1.) Refer to Figure 6-16 and Figure 6-17. Use a wrench to loosen the two retaining rings from the
switches before removing the CT position switch from the table.
2.) Remove the retaining rings from PET position switches as well.
Figure 6-16: PET Position Switches (Shown with Pedal Assembly Attached to Rail)
Socket-Head
Cap Screws
Retaining Rings