6.2.9 Completion
This section contains the instructions for the Field Engineer.
1.) Refer to Figure 6-139. If new Gantry keypads were shipped, install them at this time. (The new
keypads replace the gantry tilt buttons with PET and CT position buttons.)
Figure 6-139: Gantry Keypad
2.) Remove tags and lock.
3.) Power up the system.
4.) If necessary, push the table base into the CT detent. (The Table latch should be in the service
position at this time. (See page 184.)
5.) Press the E-Stop Reset on the gantry keypad to engage the clutch motor.
6.) Verify the Table Position Latch Arm works:
a.) Refer to Figure 6-139. Press and hold the PET position button to drive the base to the PET
detent. (The PET position LED stops flashing and remains lit.)
b.) Press and hold the CT position button to drive the base back to the CT detent.
7. Shutdown applications:
a.) Click the SERVICE icon to open the Service Desktop.
b.) Click the TOOLBOX/UTILITY icon to open the tab.
c.) Click on APPLICATIONS SHUTDOWN.
7.) Open a winterm.
8.) Login as root.
9.) Insert the Patch CD, P/N 2374180, into the CT CD ROM drive.
10.) Wait for drive busy light to turn Off.
11.) Type/enter: installpkg.
12.) Click CDROM.
13.) Click/highlight the Patch name.
14.) Click INSTALL.
15.) After the patch successfully installs, type/enter: reconfig in the winterm.
16.) Click HARDWARE to open the corresponding panel.
PET button drives table
CT button drives table
base to CT detent.
base to PET detent.
E-Stop Reset
LED lights when base
enters PET detent.