DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 136 Chapter 6 - Table
6.2.6.4 Determine the PET Detent Location
The specified distance between the CT scan plane and PET scan plane equals 675 mm
+/-14 mm. For the PET detent holes drilled into the Revision 3 rail and the PET plugs (Revision 0,
1, and 2), position the PET detent 675 mm from the CT detent. The PET detent has slots to
accommodate +/-14 mm range of tolerance. Regardless of the table rail revision, If the distance
recorded on page 113 is less than 661 mm, a new set of PET detent holes need to be drilled into
the rail, and the PET detent and PET detent adjuster shifted closer to the CT detent.
When the listed procedures are completed below, continue to Section 6.2.6.10 on page 142.
Side Rail
Revision
Distance Between Scan Planes
Follow procedure in the following
Sections:
3 675mm +/-14 mm N/A
Proceed to 6.2.6.10 on page 142.
3 651 to 661 mm 6.2.6.6 on page 139, then
6.2.6.8 on page 141.
3 641 to 651 mm 6.2.6.6 on page 139, then
6.2.6.9 on page 141.
2 675mm +/-14 mm 6.2.6.5 on page 137.
2 651 to 661mm 6.2.6.5 on page 137, then
6.2.6.7 on page 140, then
6.2.6.8 on page 141.
2 641 to 651 mm 6.2.6.5 on page 137, then
6.2.6.7 on page 140, then
6.2.6.9 on page 141.
1 675mm +/-14mm 6.2.6.5 on page 137
1 651 to 661 mm 6.2.6.5 on page 137, then
6.2.6.7 on page 140, then
6.2.6.8 on page 141.
1 641 to 651 mm 6.2.6.5 on page 137, then
6.2.6.7 on page 140, then
6.2.6.9 on page 141.
0 675 mm +/-14 mm 6.2.6.5 on page 137
0 651 to 661 mm 6.2.6.5 on page 137, then
6.2.6.7 on page 140, then
6.2.6.8 on page 141.
0 641 to 651 mm 6.2.6.5 on page 137, then
6.2.6.7 on page 140, then
6.2.6.9 on page 141.