Page 248 Chapter 8 - System QC, Calibration and Functional Tests
8.1.1.5 VQC Processing and Software Realignment
1.) Set the system into PET mode by clicking the [PET] icon on the Image Monitor desktop area.
The Scan Monitor displays the PET main screen.
2.) Click the [Full QA] button on the PET main screen.
3.) Select Volumetric QC from the drop-down menu. The Volumetric QC panel shown in
Figure 8-6 displays.
Figure 8-6: Initial Volumetric QC Panel
4.) Refer to Figure 8-7. Click [Load] to load the image sets that represent the CT and PET planes.
The Images Database Navigator displays showing the Patient List.
Figure 8-7: Images Database Navigator