GE Discovery LS - 7.1.2.1 Gantry Transport Mechanism

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 224 Chapter 7 - Gantry
7.1.2.1 Gantry Transport Mechanism
Refer to Figure 7-1. The PET and the CT Gantries are installed close to one another. In order to
enable service access for the gantries, the PET gantry is installed on a set of rollers on one side
and a linear bearing assembly on the other.
The linear bearing assembly is connected to a baseplate that is anchored to the floor. The gantry
is also connected to a mechanical driving mechanism that is designed to enable movement of the
gantry to its service position using an electric screwdriver to drive the lead screw of the gantry.
Figure 7-1: Gantry Transport Mechanism
Rollers
Roller Baseplate
Linear
Bearing
Assembly
Driving Mechanism
Cable Retractors