GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 3 - DLS System Functional Description Page 99
To shut down the CT and PET applications:
1.) From the Desktop Selection Area on the Image Monitor (right monitor), click the [Shutdown]
button (Figure 3-18). The Service Desktop displays.
2.) Click the [Toolboxes/Utilities] button. The Utilities/Tools menu displays.
Figure 3-19: Service Desktop Showing Utilities/Tools Menu
3.) From the Utilities/Tools menu, click Application Shutdown.