GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 284 Chapter 10 - Electrical Connections
Figure 10-11: PET System Power Interconnections
C
T
G
A
N
T
R
Y
PET
COMPUTER
ELECTRONICS
CABINET
PDU
MONITOR
T
A
B
L
E
CONTROL ROOM
ELECTRONICS ROOM
PDB
UP TO 100 ft
UP TO 100 ft
Power cable
to
PDB
Label both
ends #1
Label both
ends #2
Label both
ends #3
SCANNER ROOM
Power cable
for gantry
and table
Flexible conduit
and power cable
To scanner
system
disconnect
PET
TAPE

Table of Contents