GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 208 Chapter 6 - Table
4.) Mount the Back Cap on the Motor Base. Make sure that the flange of the Back Profile clips into
the Top Cap, as shown in Figure 6-137.
5.) Use a 3 mm Allen wrench to fasten the Back Cap into place with two sets of M4 hardware:
• M4 x 20 mm socket-head cap screws (SHCS), P/N 46-328417P3
• M4 lock washer, P/N 46-328432P2
• M4 flat washer, P/N 46-328430P2
Figure 6-137: Attach Back Cap to Motor Base
6.2.8 Finish the Mechanical Installation
1.) Wipe down the table covers, as needed, to remove any marks.
2.) Clean up the area.
This completes the Mechanical Install portion.