EasyManua.ls Logo

GE Discovery LS - 6.2.6.35 Install Optical Sensor (P;N 2350570)

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 177
6.2.6.35 Install Optical Sensor (P/N 2350570)
CAUTION Refer to Figure 6-93. To prevent damage to the optical sensor, temporarily fasten it into
place with one screw. After the sensor mounting bracket is adjusted and the clearance
checked, fasten the sensor into place with both screws.
Figure 6-93: Fasten Optical Sensor with One Screw
1.) Refer to Figure 6-93. Use a 3 mm Allen wrench to attach the optical sensor to the CT-PET
Position Latch with a single M3 x 10 mm screw, P/N 46-214695P7.
2.) Refer to Figure 6-93. Pivot the optical sensor away from the table rail.
3.) Slowly move the table into the PET detent.
4.) Refer to Figure 6-94. Carefully swing the optical sensor over the flag.
Figure 6-94: Swing the Optical Sensor into Place
Bend this bracket to adjust
the optical sensor with
respect to the flag plate.
Flag Plate

Table of Contents

Related product manuals