Chapter 6 - Table Page 219
6.5.3 Replacing the Secondary Base Motor
1.) Loosen the two coupling screws. (Refer to 1 in Figure 6-148 for position; screws are not
shown.)
Figure 6-148: Secondary Base Motor - General View
2.) Remove the four screws holding the motor on the motor base. (See 2 in Figure 6-149.)
3.) Refer to Figure 6-149. Push the coupling to close the gap between the coupling and the motor
flange, and push the motor to the side of the secondary base.
4.) Loosen the set screw holding the shaft. (No. 3 in Figure 6-148 and Figure 6-149), and push
the shaft into the clutch.
Figure 6-149: Motor - View 2
1
2
3
Push Coupling
3