6.2.6.21 Install Base CMD Adapter (P/N 2304737)
Refer to Figure 6-60. The CMD adapter has three connectors with captive hardware. It is difficult to
attach the existing cable with the CMD Adapter fastened onto the ETC I/F Board. It is easier to
attach the existing cable to the back of the CMD Adaptor before attaching the adaptor to the ETC
I/F Board.
Figure 6-60: CMD Connector, P/N 2304737
1.) Refer to Figure 6-61. Disconnect Signal Harness P/N 2271256, labeled I/F J7, from J7 on the
ETC I/F Board.
Figure 6-61: ETC Board and ETC I/F
2. Connect this end, labeled
1. Connect existing
3. Connect new base CMD
Cable to J1.
Adapter Cable to J2.
ETC I/F J7, to the ETC I/F Board.
ETC I/F J7