GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 2 - Discovery LS System Safety Information Page 53
STC Service Switches
Refer to Figure 2-14. The STC Service switches are located at the top of the STC on the backplane.
UP (enabled) is the normal operational position for these switches.
Figure 2-14: Location of STC Service Switches (Shown in Off Position) and LEDs
Table 2-2: STC Service Switch Descriptions
Label Description
S1 Momentary Pushbutton: Resets gantry drives enable circuit in PDU.
S2 Switch enables or disables the Axial Drive function. Default position is up
(enable).
S3 Switch enables or disables the HVDC function. Default position is up
(enable).
S4 Switch enables or disables Gantry 120 VAC function. Default position is up
(enable).
With S4 Off, 120 VAC to the gantry and table service outlets is controlled by
CB3 in the PDU only.
  
  







    



