DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 203
c.) Refer to Figure 6-130. Fit the front side covers over the long side covers.
Figure 6-130: Secondary Base with Side Covers and Roller Cover Open
CAUTION To prevent damage to the roller cover, do NOT try to attach the roller cover to the cover
bracket with the table in the service position. Do NOT move the table to the service position
with the roller cover in place.
6.) Use a 4 mm Allen wrench to reattach the roller cover to the cover bracket.
6.2.7.12 Attach the Interface Cover Between the DLS Table and Gantry
Refer to Figure 6-131. Remove Interface Cover, P/N 2331112, from the upgrade kit.
Use a 5 mm Allen wrench to fasten the Interface cover into place with two sets of M6 hardware:
M6 x 30 socket-head cap screws (SHCS), P/N 46-328417P16
M6 lock washer, P/N 46-328432P3
M6 flat washer, P/N 46-328430P3
Figure 6-131: Interface Cover
Use screwdriver or bolt to prevent the cover from retracting into the canister.
Attach Rear Side Covers first. Front Side Cover overlaps rear cover.