6.2.7.7 Attach Canister Brackets to Head End of the Secondary Base
1.) Locate and remove the Right Canister Bracket Assembly, P/N 2332626-2, and the Left
Canister Bracket Assembly, P/N 2332626, from the upgrade kit. (The canister brackets are
mirror images.)
2.) Refer to Figure 6-120. Position the L-shaped canister bracket so the free end points away from
the gantry.
3.) Use a 5 mm Allen wrench to attach the Canister Bracket Assemblies to the front end of the
table with four M10 x 20 flat-head beveled screws (P/N 2103580P22).
Figure 6-120: Canister Bracket Assembly (with and without Canister)
Bracket assembly ships with
bottom bolt threaded through hole.
Hang the roller canister off this
bolt to support its weight while
fastening it into place.
Roller Cover Flaps