GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 220 Chapter 6 - Table
5.) Refer to Figure 6-150. Remove the motor assembly.
Figure 6-150: Motor Assembly and Coupling Removed
6.) Take the new motor, insert the coupling, and push the coupling to the motor flange.
7.) Put the new motor in position next to the side of the secondary base.
8.) Refer to Figure 6-151. Using pliers, pull the shaft out of the clutch until the groove is visible,
and close the set screw.
Figure 6-151: Motor Showing Groove and Set Screw
9.) Move the motor to its operating position.
10.) Fasten the four screws connecting the motor to the motor base.
11.) Place the coupling between the two shafts and tighten the coupling screws.
Groove
Close set screw when
the groove is visible.