DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 174 Chapter 6 - Table
3.) Refer to Figure 6-89. Use a 3 mm Allen wrench to fasten the adjuster block to the side rail.
4.) Adjust the set screw until it just touches the PET detent.
5.) Finger tighten the hex nut against the block.
Figure 6-89: PET Detent Adjuster Fastened into Place
6.2.6.34 Shim the PET Detent Assembly
During Section 6.2.6.28, the PET and CT Detent/Flag assemblies were temporarily installed. Follow
the instructions in this section to determine whether to shim the detent assemblies to accommodate
for differences in the relationship between the linear bearing rail and the secondary side rail.
1.) Locate and remove the following hardware from the upgrade kit:
Part Number Description Quantity
46-214695P23 M5 x 12 mm Socket-Head Cap Screw 3
46-312577P28 M5 x 22 mm Screw 3
2355629 Shim As required (Four sent with kit.