6.2.6.17 Replace the ETC Interface Board
CAUTION To prevent damage to the system, observe all ESD precautions. Wear a grounded wrist strap
when handling the ETC I/F Boards.
1.) Refer to Figure 6-49. Locate and remove the ETC I/F Board, P/N 2229871, from the upgrade
kit.
2.) Unfasten the cables connected to the ETC I/F Board.
3.) Use a flat-blade screwdriver to unfasten the three screws that hold the board in place.
4.) Remove the new ETC I/F Board from its static protection bag and fasten it into place.
5.) Place the old ETC I/F Board into the static protection bag, and follow local guidelines to return
the board to GE.
6.) Fasten the ETC I/F cables back into place.
Figure 6-49: ETC Interface Board
ETC I/F fastens to the ETC Board with four screws.
ETC I/F
ETC Board