GE Discovery LS - Figure 6-25: Secondary Base Spreader Bar Clamped into Place; Figure 6-26: Support Casting with Wood

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 131
3.) Refer to Figure 6-25. Clamp the Side Pieces of the Spreader Bar to the secondary base rails.
(Do not remove any table anchors.)
4.) Tighten the Spreader Bar bolts.
Figure 6-25: Secondary Base Spreader Bar Clamped into Place
5.) Refer to Figure 6-26. Wedge a piece of wood under the clutch motor casting to support it while
removing the nine bolts that fasten it to the secondary base.
Figure 6-26: Support Casting with Wood
6.) If necessary, remove the floor mounted table lateral alignment assembly.