6.2.6.15 Install the Clutch Driver Unit Power Cable (P/N 2338088)
Refer to Figure 6-45. Make a mental note of the wire color before disconnecting A1 and A2 on the
ETC ASM side of the filter. The lugs on the new power cable have dual connectors, so the original
wires re-plug in place.
1.) Refer to Figure 6-45. Remove the wires from A1 and A2 of the ETC ASM side of the filter.
2.) Refer to Figure 6-44. Fasten Cable P/N 2338088 to J200 of the Clutch Driver Unit.
3.) Route the other end of P/N 2338088 to the AC Filters.
Figure 6-45: Filter Before Upgrade
4.) Refer to Figure 6-46. Plug the white connector into the right side of the filter and the black
connector into the left side of the filter.
Figure 6-46: Dual Connectors on P/N 2338088
Add wires to
Note colors.
this filter.
White Wire
Black Wire