2. Refer to Figure 6-69. Be careful when making the gantry connections. Look for the connectors
in the harness next to the lower left front quadrant of the gantry.
The cable ends of P/N 2323729 connect to previously unused cables located in the gantry
harness.
The cables are clearly marked, but a flashlight may be needed to read the labels.
Figure 6-69: Base Potentiometer E Cable, P/N 2323729
6.2.6.24 Reassemble the Secondary Base Drive Belt
1.) Refer to Figure 6-70. Use a 1.5 mm Allen wrench to loosen the four set screws (two each on
the potentiometer pulley and motor clutch pulley). This prevents damage to the potentiometer
when installing the drive belt.
Figure 6-70: Loosen Set Screws on Clutch Motor Pulley
To Gantry Side Cables
Loosen set screws
on both pulleys.