GE Discovery LS - Circuit Breakers and Switches - CPDU; Figure 2-23: CPDU Power Lamp (CPDU Top Cover Opened and Latched); Figure 2-24: CPDU Circuit Breakers; Hazard Awareness Indicators - CPDU

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 2 - Discovery LS System Safety Information Page 63
2.7.1.1 Hazard Awareness Indicators - CPDU
Refer to Figure 2-23. With the CPDU’s top cover hinged open, a small power lamp is visible. When
illuminated, this lamp indicates power is present within the CPDU.
Figure 2-23: CPDU Power Lamp (CPDU Top Cover Opened and Latched)
2.7.1.2 Protected Service Outlets - CPDU
Refer to Figure 2-22. The service outlet is protected by a circuit breaker. The outlet is located on
the A4 panel.
2.7.1.3 Circuit Breakers and Switches - CPDU
Circuit Breakers
Refer to Figure 2-24. There are three (3) groups of circuit breakers in the CPDU used to protect
various parts of the system.
CB1 - Console AC Power
CB3 - Table and Gantry AC Service Outlets
CB4 - Table and Gantry AC to Stationary Electronics
CB5 - Gantry Rotating Power, Tilt Power and Communications
CB6 - Main Axial Drive Power
CB7 - Master 120/208 VAC Power (CB1, 3, 4 and 5)
Figure 2-24: CPDU Circuit Breakers