DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 200 Chapter 6 - Table
8.) To close the cover:
a.) Slide the roller cover toward the table while you slide the table forward.
b.) Use a 4 mm Allen wrench to fasten the cover into place with three M6 x 12 flat-head
screws, P/N 2103580-12.
9.) Refer to Figure 6-127. Check roller cover clearances. Readjust the idler roller and cover
bracket if the cover fails to meet the following conditions:
Cover rests on the lower guide.
Cover does not rub against the top guide when pushing the table between the CT and PET
detents.
An even gap exists between the top guide and the cover.
10.) If the cover appears off center or it rubs one of the rails, readjust the cover canister.
Figure 6-127: Check Cover Clearances
11.) Once adjusted, detach the roller cover, and insert the screwdriver until finished installing the