GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 200 Chapter 6 - Table
8.) To close the cover:
a.) Slide the roller cover toward the table while you slide the table forward.
b.) Use a 4 mm Allen wrench to fasten the cover into place with three M6 x 12 flat-head
screws, P/N 2103580-12.
9.) Refer to Figure 6-127. Check roller cover clearances. Readjust the idler roller and cover
bracket if the cover fails to meet the following conditions:
• Cover rests on the lower guide.
• Cover does not rub against the top guide when pushing the table between the CT and PET
detents.
• An even gap exists between the top guide and the cover.
10.) If the cover appears off center or it rubs one of the rails, readjust the cover canister.
Figure 6-127: Check Cover Clearances
11.) Once adjusted, detach the roller cover, and insert the screwdriver until finished installing the
secondary base side covers.