6.2.7.15    Attach the Back Cap to the Foot End of the Secondary Base
Follow the procedure in this section to attach the back cover brackets and the Back Cap to the foot 
end of the secondary base. The back cover brackets ship with the cover hardware.
1.) Refer to Figure 6-134 and Figure 6-135. Locate and remove the two back cover brackets, P/N 
2323712, from the cover hardware bag.
NOTICE For best results, dry fit the Back Cap before tightening the back bracket hardware into place. 
Leave the hardware loose enough to shift the bracket, but tight enough to hold it in place. 
Fit the Back Cap into place, and insert a skinny screwdriver blade or thick wire through each 
set of Back Cap and bracket holes. Shift the bracket to align its holes to the Back Cap. 
Expect to take multiple tries to fit the Back Cap to the table.
Figure 6-135:  Attach the Back Cover Brackets to the Secondary Base
2.) Use an 8 mm Allen wrench to fasten each bracket into place with two sets of the following M10 
hardware:
• M10 x 20MM socket head cap screws (SHCS), P/N 46-328417P22
• M10 lock washer, P/N 46-328432P4
• M10 flat washer, P/N 46-328430P4