EasyManua.ls Logo

GE Discovery LS - 6.2.7.14 Attach the Top Cap Assembly to the Secondary Base

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 205
6.2.7.14 Attach the Top Cap Assembly to the Secondary Base
Refer to Figure 6-134. The Top Cap Assembly, P/N 2320161, attaches to the foot end of the
secondary base, and rests on flanges attached to the side rails. When the table moves back to the
CT detent, the metal Table Top Cover, P/N 2333078, slides beneath the Top Cap Assembly.
The Top Cap attaches with a bracket to the foot end of the secondary base.
1.) Use a 5 mm Allen wrench to fasten the Top Cap bracket into place with two sets of the following
M5 hardware:
M5 x 16 socket head cap screw (SHCS), P/N 46-328417P505
M5 lock washer, P/N 46-328432P8
M5 flat washer, P/N 2236649
Figure 6-134: Table Top Cap Assembly (P/N 2320161)
Use 5 mm Allen wrench.
Use 8mm Allen wrench.

Table of Contents

Related product manuals