6.2.6.5    Attach the PET Template and Drill the Standard PET Detent Holes
Revision 0, 1 and 2 tables use the PET Drill template. Revision 3 tables do not use this template. 
Follow the instructions in this section to attach the PET Drill Template,
P/N 2355334, to the right secondary base side rail.
1.) Locate and remove the following hardware from the upgrade kit:
2.) Refer to Figure 6-32. Insert the PET Drill Template into the fifth access hole, counting from the 
gantry end of the right secondary base rail.
3.) Rest the edges of the raised portion of the template against the bottom and front edge (gantry 
side) of the access hole.
Figure 6-32:  PET Drill Template 
Part Number Description Quantity
2355225 PET Drill Template 1
2355559 PET Drill Template Clamp Block 2
46-328417P19 M6 x 50 mm Socket-Head Cap Screw 4
2315373-4 M6 Lock Washer 4
2109878-3 M6 Flat Washer 4
Interior View
Exterior View
Push template against the front edge of the access hole (toward the gantry). 
 
Bottom edge rests against
bottom edge of 5th access hole.
This edge 
rests against
front edge of
5th access hole.
Clamp Block