6.2.6.41    Install the Fourth Top Cover Bracket
1.) Locate and remove the following hardware from the installation kit:
2.) Refer to Figure 6-109. Use a 3 mm Allen wrench to fasten the fourth Top Cover Bracket, P/N 
2332371
, to the back of the CT-PET Position Latch with the M4 hardware.
Figure 6-109:  Install Top Cover Bracket, P/N 233371
6.2.6.42    Dress the Rest of the Cables and Close the ETC
1.) Dress the cables installed between the rear of the table bulkhead and the ETC, and tie-wrap 
to the existing table harness as needed.
2.) Dress and tie-wrap the cables in the table retractor area as needed.
3.) Dress (twist and tuck) the CT-PET Position Latch cables to keep them out of the path of travel.
4.) Close the ETC Assembly:
a.) Make sure all connections were made to the ETC and ETC I/F.
b.) Fasten any cable connector hardware.
c.) Make sure all cables between the ETC and CT-PET Position Latch are secure and tie-
wrapped out of the way of the ETC cover supports and mounting tray.
d.) If necessary, fasten the ETC mounting tray back into place.
Part Number Description Quantity
2332371 Cover Bracket (Right Rear) 1
46-328417P4 M4 x 12 mm Socket-Head Cap Screw 2
46-328432P2 M4 Lock Washer 2
46-328430P2 M4 Flat Washer 2
Top Cover Bracket
CT-PET Position Latch