DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 123
6.2.5.2 Unfasten and Remove the Cables
The currently installed ground wire may be reused. Reconnect the new Clutch Control and Base
Axis Motor cables to the original gantry cables, and fasten the new Base Potentiometer cable to a
different set of gantry cable connectors.
1.) Refer to Figure 6-14. Disconnect the three cables and ground wire from the KB Box.
2.) Cut any tie-wraps, as needed, to remove the three motor clutch cables and ground wire from
the table.
3.) Follow the cables to the front of the table, and disconnect them from the corresponding gantry
cables. Remove the ground wire from the ground bus bar.
4.) Follow local guidelines to discard the cables and ground wire.
Figure 6-14: KB Box and Clutch Motor Assembly
6.2.5.3 Unfasten and Remove the KB Box
1.) Refer to Figure 6-14. Use a 3 mm Allen wrench to remove the three socket-head cap screws
that fasten the KB box to the bracket on the casting.
2.) Return the KB Box (and clutch motor casting) to GE Headquarters.
Disconnect three cables
and ground wire.
Use 3 mm Allen wrench to
remove 3 socket-head
cap screws.