6.2.6.2 Determine the Right Secondary Base Rail Revision
Determine the secondary base revision by counting the number of access holes in the right rail
while facing the Gantry. All tables use the CT Drill Template.
Seven rail access holes: Revision 0 table needs the PET Plug and CT Plug. Drill four PET
holes and nine CT holes.
Five rail access holes: Revision 1 or 2 table needs the PET Plug. Drill four PET holes and
six CT holes.
Four rail access holes: For Revision 3 table, drill one CT hole. (May need to drill four PET
holes if the distance between the CT scan plane and PET scan plane falls outside the
specified range.)
Figure 6-28: Determine the Right Secondary Base Rail Revision
Revision 0 rail has seven access holes.
Access Hole
Revision 1 rail has five access holes.
Revision 2 rail has five access holes.
Revision 3 rail has four access holes.