Page 184 Chapter 6 - Table
5.) Refer to Figure 6-105. Use a 2.5 mm Allen wrench to loosen the solenoid spring bracket.
Figure 6-105: Adjust the Solenoid Spring
6.) Refer to Figure 6-104 and Figure 6-105. Lift the CT-PET Position solenoid latch arm. (Let the
tension on the spring pull the bracket toward the solenoid.)
7.) Pull the bracket 3 mm away from its current position, and tighten the bracket screws using a
2.5 mm Allen wrench. (Maintain spring tension until both screws are tightened.)
8.) Refer to Figure 6-104. Depress and release solenoid latch arm with your fingers. Make sure
the latch arm returns to the initial (upward) position.
9.) Move the table forward and back into position to verify the adjustment. (The arm should clear
the service latch.)
CAUTION To prevent damage to the CT-PET Position assembly, leave the Latch in the service (raised)
position until you complete the power up and characterization procedures.
10.) Refer to Figure 6-106. Raise the Table Latch to the service position, and leave it in the raised
position until the table is characterized.
Figure 6-106: Latch in Service Position
Screws on Solenoid
Lift up latch arm.
Spring Bracket