NOTICE The captive screw at the bottom of the ETC mounting tray is rounded to fit easily into the
connector. Pivot the ETC mounting tray into place, and push down until the captive screw
drops into place. Use a flat-blade screwdriver to fasten it in place.
e.) Check along the bottom of the mounting tray, and reposition any wires and cables to
prevent them from becoming pinched beneath the mounting tray.
f.) Fasten the ETC cover into place with the captive hardware.
* Finger tighten the knurled screws; do not use a screwdriver.
* Make sure all the ETC cables and connectors clear the cover.The cover comes off to
set the flags.
5.) Use a 1/8 inch Allen wrench to fasten the braided ground and the two screws that fasten the
right and left covers to the table elevator (scissor section).
6.) Move the table jumpers to the storage position, then reattach the Table Side Covers.
7.) Check the Table Latch one more time. Make sure it is pinned in the vertical (service) position.
6.2.6.43 Install the Cosmetic Cradle Caps
1.) Refer to Figure 6-110. Locate and remove the 10 black plastic cosmetic cradle caps, P/N 46-
221523P10, from the upgrade kit.
2.) Insert a cosmetic cap into each of the ten holes located in the foot end of the cradle.
3.) Press down on the cap with your thumb until it snaps into place.
4.) If necessary, gently tap the caps with a rubber mallet to snap them into place.
Figure 6-110: Install Ten Cosmetic Cradle Caps, P/N 46-221523P10
Cosmetic caps inserted into holes
in foot end of the cradle top.
Five Cosmetic Caps Per Side