GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 197
6.2.7.9 Attach the Roller Cover Canister to the Secondary Base
For best results, use two people to install the Roller Cover Canister, P/N 2331848.
Use a 5 mm Allen wrench to fasten the canister assembly to the table with four sets of the following
M6 hardware:
M6 x 16 socket-head cap screw, P/N 46-328417P13
M6 flat washer, P/N 46-328430P3
M6 lock washer, P/N 46-328432P3
1.) Refer to Figure 6-121. Flip the canister assembly forward, so the drip tray and cover face the
ceiling. Carefully lower the canister assembly between the gantry and table.
2.) Refer to Figure 6-120. Each canister bracket assembly arrives with two of the M6 screws
fastened through the bottom of the bracket. Slide the lower set of canister holes over the
screws to support the weight while aligning and fastening the assembly into place.
3.) Center the canister to the secondary base cover guides, and check for levelness before
fastening it into place.
Figure 6-122: Roller Canister in Place (P/N 2331848)
Each bracket arrives with bottom screw in place.
M6 Hardware