6.2.6.30 If a CT Plug Is Required
1.) Refer to Figure 6-83. If working with a Revision 0 rail, the CT Plug, P/N 2342039, is installed
in the seventh access hole from the front of the table.
Figure 6-83: Revisions 0, 1, and 2 Right Rails
2.) Use the following hardware to install the CT Plug into the seventh hole in the rail:
a.) Refer to Figure 6-84. Insert the plug from the inside of the table.
b.) Fasten the nuts and washers to the rail exterior and screws to the rail interior.
Figure 6-84: CT Plug
Part Number Description Quantity
2342039 CT Position Plug 1
46-312577P28 M5 x 25 mm Socket-Head Cap Screw 5
2236649 M5 Flat Washer 5
46-328432P8 M5 Lock Washer 5
46-328425P11 M5 Hex Nut 5
Revision 0 Side Rail (Seven Access Holes)
CT Plug in seventh access hole.
Exterior View
Interior (Inside the Table)
Mounting Holes
Mounting Holes