Page 120 Chapter 6 - Table
c.) Refer to Figure 6-11.
* If the Calibration Bar has a metal tab on the left side of the bar, proceed to Step d.
* If the Calibration Bar does not have a metal tab on the left side of the bar, proceed to
Section 6.2.4.3.
d.) Use a 5 mm Allen wrench to remove the two M6 socket-head cap screws that fasten the
calibration bar into place.
e.) Refer to Figure 6-11. Trim the tab flush with the left side of the Calibration Bar.
Recommended: If a vice is not available, use one of the C-clamps to clamp the Calibration Bar to
the end of a secondary base side rail. Use a hacksaw to cut the tab.
Figure 6-11: Remove the Tab from the Calibration Bar
If Calibration Bar has a
tab at this end, remove it.
Calibration Bar should
look like this.