6.2.5.9 Remove the INTF SW and CL CONT I Cable (P/N 2309066) from the Table
Some versions of the M1 Table do not have Table bulkheads. The cables attach directly to the
cable connections in the table retractor. If a Table bulkhead is not present, the new bulkhead (P/N
2309772) that comes with the upgrade kit will be installed during Section 6.2.6.11.
1.) Refer to Figure 6-21. Locate and detach the 25 pin D-Type connector and pigtail from J7 of the
ETC I/F Board. (If J7 cannot be located on the ETC I/F Board, see Figure 6-19 on page 126.
2.) Follow the cable to the rear connection of J12 of the Table bulkhead. Cut tie-wraps and cable,
as needed, to remove P/N 2309066 and the pigtail from the table.
3.) Temporarily disconnect the cable from J12 on the front of the Table bulkhead.
4.) Use a 5 mm Allen wrench to unfasten the Table bulkhead hardware to access and detach the
cable connectors from the Table bulkhead.
5.) Remove Cable P/N 2309066 from J12 at the rear of the Table bulkhead and discard.
Figure 6-21: Disconnect J7