EasyManuals Logo

GE Discovery LS User Manual

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 151
6.2.6.19 Install the CT-PET Position Optical Sensor (P/N 2350570)
For best results, route P/N 2350570 under the ETC Board before making the connection.
1.) Locate and remove the CT-PET Position Optical Sensor from the kit.
2.) Refer to Figure 6-52. Trim about 1 inch of the heat shrink insulation from the connector. (The
full length of insulation may interfere with the ETC Cover.)
Figure 6-52: Trim Insulation from P/N 2350570 Connector
3.) Refer to Figure 6-53. Carefully route the sensor across the ETC assembly and through the
slot in the side bracket.
Figure 6-53: Route Optical Sensor Across ETC Board
4.) Fasten the D-Type connector to J8 of the ETC I/F Board.
5.) Locate the old J8 connector detached and taped off in Section 6.2.5.7.
Cut insulation back
by about 1 inch.
Route the Optical Sensor through the slot in the side bracket.
Be careful, Optical
Sensor is very fragile.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Discovery LS and is the answer not in the manual?

GE Discovery LS Specifications

General IconGeneral
BrandGE
ModelDiscovery LS
CategoryMedical Equipment
LanguageEnglish

Related product manuals