DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 151
6.2.6.19 Install the CT-PET Position Optical Sensor (P/N 2350570)
For best results, route P/N 2350570 under the ETC Board before making the connection.
1.) Locate and remove the CT-PET Position Optical Sensor from the kit.
2.) Refer to Figure 6-52. Trim about 1 inch of the heat shrink insulation from the connector. (The
full length of insulation may interfere with the ETC Cover.)
Figure 6-52: Trim Insulation from P/N 2350570 Connector
3.) Refer to Figure 6-53. Carefully route the sensor across the ETC assembly and through the
slot in the side bracket.
Figure 6-53: Route Optical Sensor Across ETC Board
4.) Fasten the D-Type connector to J8 of the ETC I/F Board.
5.) Locate the old J8 connector detached and taped off in Section 6.2.5.7.
Cut insulation back
by about 1 inch.
Route the Optical Sensor through the slot in the side bracket.
Be careful, Optical
Sensor is very fragile.