4.) Refer to Figure 6-73. Position the belt so the marks on the belt line up with the edges of the
belt clamp. (Orient the belt clamp with the smooth surface facing up and the corrugated side
against the teeth on the belt.)
5.) Use a 4 mm Allen wrench to fasten the clamp into place with four sets of belt clamp hardware.
Figure 6-73: Reassemble the Drive Belt
3. Refer to Figure 6-72. If the Drive Belt Idler Pulley was loosened, push the pulley into its
perpendicular position with a large screwdriver, reinsert the hardware, and use a
5 mm Allen wrench to tighten the pulley screws.
Marks on the belt align
with edges of belt clamp
(corrugated side down).