EasyManua.ls Logo

GE Discovery LS - 12.4.2 Base Axis Does Not Move

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 12 - Troubleshooting Page 333
12.4.2 Base Axis Does Not Move
If the base axis does not move, perform the following checks:
Figure 12-12: Troubleshooting for Base Axis Does Not Move
Base Axis does not move
Is E-STOP circuitry
activated or Cradle Release
button pressed?
Reset the E-STOP
circuitry or press the
Cradle Release button
yes no
Does a mechanical interference exist?
Check the locking mechanism including the
solenoid functionality, use the PET position Jog
button on the Control Panel
yes
no
Release the
mechanical
interference
Check:
1. The path from the ELEVATION/TILT(BASE Axis ) Amplifier to
the motor include the Amplifier itself
PNs:
1. 2323724
2. 2309058
3. 2323720
4. 2323722
End of test

Table of Contents