12.4.2 Base Axis Does Not Move
If the base axis does not move, perform the following checks:
Figure 12-12: Troubleshooting for Base Axis Does Not Move
Base Axis does not move
Is E-STOP circuitry
activated or Cradle Release
button pressed?
Reset the E-STOP
circuitry or press the
Cradle Release button
yes no
Does a mechanical interference exist?
Check the locking mechanism including the
solenoid functionality, use the PET position Jog
button on the Control Panel
yes
no
Release the
mechanical
interference
Check:
1. The path from the ELEVATION/TILT(BASE Axis ) Amplifier to
the motor include the Amplifier itself
PNs:
1. 2323724
2. 2309058
3. 2323720
4. 2323722