a.) Attach Base Potentiometer E, P/N 2323729, to J1 of the bulkhead. (Refer to Figure 6-69.
Cable P/N 2323729 has a pigtail permanently attached to the gantry end of the cable.)
b.) Attach Clutch Control E, P/N 2323726, to J2 of the bulkhead.
c.) Attach Base Axis Motor E, P/N 23237720, to J3 of the bulkhead.
3.) Locate and remove Ground Cable, P/N 2340388, from the kit.
4.) Refer to Figure 6-67. Attach the J3 ground and P/N 23237720 to the bulkhead with the
following hardware from the kit:
1. Refer to Figure 6-68. Tie-wrap the cables and ground wire into place along the secondary
base rail, then route them under the front of the table to make the connections to the original
motor clutch cables in the gantry. (See page 123 for the clutch motor cable removal
instructions.)
Figure 6-68: Route Clutch Motor Cables Beneath Front of Table
Part Number Description Quantity
46-312577P27 M5 x 20 mm Socket-Head Cap Screw 1
2236649 M5 Flat Washer 2
46-328432P8 M5 Lock Washer 1
46-328425P11 M5 Hex Nut 1
Tie-wrap cables along side rail.
Cables routed beneath front of table.