EasyManua.ls Logo

GE Discovery LS - Figure 6-71: Drive Belt Fastener Plate; Figure 6-72: Drive Belt Idler Pulley

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 162 Chapter 6 - Table
2.) Refer to Figure 6-71. Wrap the belt around the two drive pulleys, and bring the ends together
over the fastener plate.
Figure 6-71: Drive Belt Fastener Plate
3.) Optional: Loosen the idler pulley to release tension on the belt.
a.) Refer to Figure 6-72. Use a 5 mm Allen wrench to remove the top two and bottom left
socket-head cap screws that fasten the drive belt idler pulley to the side rail.
b.) Loosen the lower right screw, but leave it in place.
Figure 6-72: Drive Belt Idler Pulley
Clamp belt into place
over the fastener plate
(shown with clamp
hardware).
Four Socket-Head Cap Screws (Right lower screw left in place.)
Drive Belt Idler Pulley

Table of Contents