EasyManua.ls Logo

GE Discovery LS - Figure 6-9: Characterization Calibration Bracket; Figure 6-10: Calibration Bar with Tab

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 119
3.) Refer to Figure 6-9. Remove the characterization bracket, located near the fans on the right
front corner of the ETC chassis.
a.) Use a 5 mm Allen wrench to remove the M6 socket-head cap screw that fastens the
characterization calibration bracket into place. (Bracket is silk-screened: LOWER LIMIT
SWITCH and LOWER CHAR POINT.)
b.) Discard the bracket.
Figure 6-9: Characterization Calibration Bracket
4.) If necessary, modify the Calibration Bar:
a.) Refer to Figure 6-10. Locate the Calibration Bar fastened to the right side of the table
casting, beneath the ETC assembly.
b.) Fasten the bar back into place with the original M6 hardware.
Figure 6-10: Calibration Bar with Tab
Calibration Bracket
Remove this tab (interferes with the
CT-PET Latch Assembly attaches
to this bearing block.
CT-PET Latch Assembly).

Table of Contents

Related product manuals