DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 182 Chapter 6 - Table
6.2.6.38 Adjust the Table Latch
Follow the procedure in this section to adjust the CT-PET Position assembly and Table Latch.
1.) Push the table into the CT position. Lower the Table Latch bar and pin into place.
(Figure 6-103 shows the Table Latch pin in place.)
2.) Refer to Figure 6-101 and Figure 6-102. Use a 3 mm Allen wrench to adjust the screw to lower
the solenoid latch arm and create a gap of about 3 to 3.5 mm between the bottom of the detent
and the bottom of the Table Latch Arm. (The latch should not allow forward or backward table
movement at this time.)
Figure 6-101: Adjust CT Position Assembly to Clear Table Latch Arm
Figure 6-102: Correctly Adjusted Position Latch Arm
3 mm Allen Wrench
Solenoid
Latch Arm
Bottom of Table latch should clear the detent by 3 mm.
Use the 3 mm Allen wrench as a gauge for 3 to 3.5 mm gap here.