DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 76 Chapter 2 - Discovery LS System Safety Information
• Refer to Figure 2-30. The PET Gantry E-Stop button is located on the back of the gantry
close to the back of the gantry.
Figure 2-30: PET Gantry E-Stop
2.9.2 CT-PET Table Safety
DANGER • NEVER PUT YOUR ARMS OR LEGS BETWEEN THE SECONDARY BASE
AND THE SUPPORTING ARMS OF THE ELEVATION MECHANISM AS THEY
COULD BE TRAPPED WHEN THE TABLE DESCENDS.
• NEVER STAND OR PUT YOUR HANDS INSIDE THE SECONDARY TABLE
BASE IF THE TABLE IS NOT LOCKED INTO ONE OF ITS END POSITIONS.
WARNING DO NOT ATTEMPT TO MOVE THE TABLE TO ITS HOME POSITION WHEN IT IS IN THE PET
SCAN POSITION OR TRAVELING BETWEEN THE PET AND CT SCAN POSITIONS. THE
TABLE MUST ONLY BE MOVED TO HOME POSITION FROM THE CT SCAN POSITION.
FAILURE TO COMPLY WITH THIS WARNING MAY RESULT IN THE STRETCHER COLLIDING
WITH THE CT GANTRY WITH CONSEQUENT DAMAGE TO THE PATIENT AND/OR THE
TABLE.