GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 159
4.) Refer to Figure 6-66. Attach three tie-wrap mounts, P/N 46-208747P2, to the front end of the
table rail. (If possible, align each adhesive mount to a pre-drilled and tapped M5 hole.)
5.) Orient the mounts vertically as shown in Figure 6-65, and wipe the area with alcohol to ensure
a good grip.
6.) Use M5 screws to fasten the three mounts to the rail at the front of the (Revision 3) table.
Note: If the rail does not have pre-drilled mounting holes (Revision 0, 1, and 2 tables), use the adhesive
mounts without the M5 screws.
Figure 6-66: Attach Adhesive Mounts to Front of Table Rail
6.2.6.23 Attach Clutch Motor Cables
1.) Locate and remove the following cables from the upgrade kit:
2.) Refer to Figure 6-67. Attach the cables to the Clutch Motor bulkhead bracket. The bulkhead
bracket fastens to the Clutch Motor casting.
Figure 6-67: Clutch Motor Cables
Part Number Description
2323729 Base Potentiometer E Cable
2323726 Clutch Control E Cable
2323720 Base Axis Motor E
Clutch Motor Bulkhead
Bulkhead Ground
Ground from J3 Connector