6.2.7.13 Fasten the Rear Covers to the Secondary Base
The Top Cover attaches directly to the table, and travels with the table as it moves between the PET
and CT detents. The cover brackets were installed in Section 6.2.6.40 and 6.2.6.41. Follow the
procedure in this section to fasten the Top Cover to the table brackets.
1.) Locate and remove the metal Table Top Cover, P/N 2333078, from the kit.
Refer to Figure 6-133. The Table Top Cover has two extensions attached by metal strips to
the front edge of the cover. The metal strips flex so the extensions can fill the gap between
the elevation drive assembly and the side covers
The front edge also has a notch to accommodate the elevation drive assembly. Be careful
not to catch the extensions or bend the metal strips while maneuvering the Top Cover into
place.
2.) Refer to Figure 6-132. Square the Table Top cover between the side covers, leaving an even
gap on both sides.
Figure 6-132: Square the Table Top Cover
3.) Refer to Figure 6-133. Use a 2.5 mm Allen wrench to fasten the Table Top Cover into place
with four sets of the following M4 hardware:
M4 x 10 mm stainless steel button head screw, P/N 2109872-2
M4 lock washer, P/N 46-328432P2
M4 flat washer, P/N 46-328432P2
Figure 6-133: Top Cover Fastened into Place
Even the gap between top and side covers.
Cover extensions fastened to Top Cover by metal strips.