EasyManua.ls Logo

GE Discovery LS - Page 351

GE Discovery LS
358 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Appendix B - System Calibration Quick Guide Page 351
3. CTC 10 - 15 Minutes
1. PET
2. Full QA
3. Coincidence Timing
4. Acquisition
5. Enter acquisition time = 5 min.
6. Accept Setup
7. Start scan
8. Wait until acquisition is completed.
9. Full QA
10. Coincidence Timing
11. Processing
12. Select
13. Select acquisition data by date and time.
14. OK
15. Calculate
16. OK
To view the progress of the calibration in the
Error Log File, proceed as follows:
1. Utilities
2. Error Log Viewer
3. Message Level = Programmer
4. Update
Error messages will also display in a message
box that opens during the process. In case of
failure, call Service.

Table of Contents

Related product manuals