GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Page 354 Appendix B - System Calibration Quick Guide
8. Clean Database 20 Minutes
CT Network Queue:
1. Image Works
2. Network
3. Select Remote Host
4. Ping Dicom Host
5. Queue
6. Network
7. Delete old exams or images.
CT Recon Queue:
1. Recon Management
2. Restart Queue
CT Archive:
1. Image Works
2. Queue
3. Archive
4. Scroll through images.
5. Delete old exams or images.
PET Data:
1. Utilities
2. Delete
3. Type of Data = Patient
4. Sort by = Date-Time
5. Include = All
6. Select all patients except from the last week.
7. Select
8. Delete Selection.
9. Delete Items.
This process deletes all data from memory
except for data from the last week. It is
recommended to back up all data before
clean-up.
If not able to perform a backup before
the clean-up, perform a partial clean-up
by deleting all CT data entered in the
PET database and for which records
still exist in the CT database.