EasyManua.ls Logo

MacDon D1 Series - Calibrating Feeder House - John Deere S7 Series

MacDon D1 Series
546 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
215184 456 Revision A
1022782
A
B
Figure 9.337: John Deere S7 Display Checking Sensor
Voltage
8. Select SENSOR tab (A) to view sensor voltages. Center
header height sensor voltage (B) must be between 0.5 and
4.5 V, with at least 3 V of variation between 0 and 4 on the
float indicator box.
NOTE:
If the optional Auto Lateral Tilt AHHC kit is installed, the left
and right header height sensors must also be in the same
0.54.5 V range.
9. If sensor voltage adjustment is required, refer to Adjusting
Voltage Limits One-Sensor System, page 314.
Calibrating Feeder House John Deere S7 Series
Feeder house calibration must be done before header calibration.
For best performance of auto header height control (AHHC), perform these procedures with center-link set to D. When
setup and calibration are complete, adjust center-link back to desired header angle. For instructions, refer to the header
operators manual.
NOTE:
Changes may have been made to the combine controls or display since this document was published. Refer to the combine
operators manual for updates.
1. Ensure center-link is set to D.
2. Rest header on down stops and unlock float.
1022778
A
Figure 9.338: John Deere S7 Display Harvesting Page
3. On the HARVESTING page, select MENU icon (A) in the
bottom right corner of screen. The MENU opens.
SETTING UP AUTO HEADER HEIGHT CONTROL

Table of Contents

Related product manuals