The 38DL PLUS simultaneously displays a thickness readout and an A-scan view for
waveform verification. The microprocessor of the 38DL PLUS continuously adjusts
the receiver setup so that every measurement is optimized for reliability, range,
sensitivity, and accuracy. An advanced internal datalogger can store up to
475000 thickness measurements and 20000 waveforms.
The 38DL PLUS operates with a full line of single element and dual element
transducers to measure material thicknesses between 0.08 mm and 635 mm (0.003 in.
and 25 in.). The temperature range of measured materials can be between 20 °C and
500 °C (4 °F and 932 °F) depending on the material characteristics, the transducer,
and the measurement mode. You can also use single element or dual element
transducers for echo-to-echo measurements.
You can connect the 38DL PLUS to a printer and to a computer using the bidirectional
serial USB/RS-232 communication ports.
Advanced Measurement Features
THRU-COAT measurement
Temperature-compensated measurement
•Min/Average mode
•EMAT transducer capability
Measurement-related status flags and alarms
Full VGA color transflective LED back-lite display
Automatic probe recognition for the standard D79X and MTD705 series
transducers
Dynamic default gain optimization
V-path calibration to build custom V-path correction tables for any dual element
transducer
Warning against calibration doubling
Calibration for unknown material sound velocity and/or transducer zero
Echo-to-echo measurements
Fast scan mode with 30 readings per second
Manual gain adjustment in 1-dB steps
Hold or blank thickness display during loss-of-signal (LOS) conditions
Hold minimum, maximum, or both minimum and maximum functions
Differential thickness display relative to the set point in absolute or percentage
ratio

Table of Contents

Other manuals for Olympus 38DL PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus 38DL PLUS and is the answer not in the manual?

Olympus 38DL PLUS Specifications

General IconGeneral
Operating Temperature-10 °C to 50 °C (14 °F to 122 °F)
Data LoggerYes
InterfaceUSB, RS-232
Thickness Range0.08 mm to 635 mm (0.003 in. to 25 in.) depending on material and transducer
DisplayBacklit LCD
Battery TypeThree alkaline or NiMH AA cells
PulserAdjustable square wave pulser
Receiver Bandwidth0.5 MHz to 20 MHz
LanguagesEnglish, French, German, Spanish, Italian, Russian, Japanese, Chinese
Resolution0.001 mm or 0.0001 in.
Transducer Frequency2 MHz to 30 MHz
MaterialPlastic housing