List and sweep mode
R&S
®
SMA100B
208User Manual 1178.3834.02 ─ 09
The remote commands required to define the list mode data are described in
Chapter 14.16.7, "SOURce:LIST subsystem", on page 665.
Access to "Edit User Correction Data":
► "Level" > "User Correction" > "Edit User Cor. Data"
The editor for user correction data provides a table with RF frequency and power
values and standard navigation functions.
The remote commands required to define the user correction data are described in
Chapter 14.16.3, "SOURce:CORRection subsystem", on page 634.
All columns in a row must contain values. Cells with missing values are therefore filled
automatically, using the value of the previous row.
If you use global dwell time in list mode, consider also that the instrument uses the
value set with Global Dwell Time for all list steps and not the values from the list.
List editor