Remote control commands
R&S
®
SMA100B
466User Manual 1178.3834.02 ─ 09
:MMEMory:RDIRectory <Directory>
Removes an existing directory from the mass memory storage system. If no directory
is specified, the subdirectory with the specified name is deleted in the default directory.
Setting parameters:
<Directory> string
String parameter to specify the directory to be deleted.
Usage: Event
:MMEMory:STORe:STATe
<savrcl_state_nr>, <file_name>
Stores the current instrument setting in the specified file.
The instrument setting must first be stored in an internal memory with the same num-
ber using the common command *SAV.
Setting parameters:
<savrcl_state_nr> Corresponds to the specific <number> defined with the *SAV
command, e.g. *SAV 4.
<file_name> String parameter to specify the file name with extension
*.savrcltxt.
Usage: Event
Manual operation: See "Save" on page 319
:MEMory:HFRee?
Returns the used and available memory in Kb.
Return values:
<TotalPhysMemKb> integer
Total physical memory.
<ApplicMemKb> integer
Application memory.
<HeapUsedKb> integer
Used heap memory.
<HeapAvailableKb> integer
Available heap memory.
Usage: Query only
MMEMory subsystem